X-RAD 320 Series

Biological Irradiator

Precision X-Ray, PXi, X-rad 320 x-ray system for delivering a precise radiation dosage to specimens in a biological or small animal research lab.

Overview

The X-RAD 320 is a self-contained X-ray system for delivering a precise radiation dosage to specimens in a biological or small animal research laboratory. 

The shielded cabinet includes an Adjustable Specimen Shelf, Sample Viewing Window and Beam Hardening Filter Holder.

The TouchRAD Control Panel is a multi-user, password protected touch screen interface with a transportable database that can track individual system usage.

Passwords, programmed exposure settings, and database management are controlled by a super-user.

The X-ray tube is used specifically for radiation therapy having a highly homogenous beam. 

Features

Available Options:

Specifications

Cabinet Features
Adjustable Sample Shelf: 20cm to 90cm SSD
Changeable Beam Conditioning Filter Slides, (2) supplied
User Entry Port to introduce small tubing and cables into chamber area
Complies with Domestic and International Regulations for safety of cabinet X-ray systems

Cabinet Size and Weight
Overall Dimensions: 78” (195cm)h x 38” (95cm)w x x42” (105cm)d
Irradiation Chamber: 41” (102cm)h x 30” (75cm)w x 34.5” (86cm)d
Weight: 5500 lbs (2500 kg)

Power Requirement
1N PE 230 VAC ±5%, or 3N 230 VAC ±10%, 50/60 Hz, 12.5 KVA

High Voltage Generator
Maximum Output Voltage: 450 kV (tube limited to 320 kV)
Maximum mA: 45

X-ray Tube
Maximum Potential: 320 KV
Maximum Power: 4000 W
Type: Metal Ceramic, Fixed Anode, Oil Cooled
Focal Spot: 8 mm2 (per EN12543)

Dose Output
3 Gy/min at 320KV, 12.5mA, 50cm SSD, (HVL≈1mm Cu)
1 Gy/min at 320KV, 12.5mA, 50cm SSD, (HVL≈4mm Cu)
>15 Gy/min at 320KV, 12.5mA, 50cm SSD, (No Beam Hardening))

Operators Control
KV Setting & Display Accuracy: 5KV – 320 KV in 0.1 KV increments
mA Setting & Display Accuracy: 0.5mA to 45 mA in 0.01 mA increments
Settings Accuracy: < 1%
Exposure Timer: 1-9999 seconds
Programmable Settings: >1,000 locations to recall exposure parameters
Password Protected User Accounts (> 1000 users)
Ability to assign users access to only specific programs.
Excel® Database transportable via USB Flash memory

Precision X-Ray, PXi, X-RAD 320, precise radiation dosage to specimens or small animal research

Overview

The X-RAD 320Dx is a self-contained X-ray system for delivering a precise radiation dosage to specimens in a biological or small animal research laboratory. 

This advanced version of the X-RAD 320 allows researchers to perform irradiation procedures by rotating the X-ray source vertically downward or horizontally for additional capability. 

The larger shielded cabinet includes an Adjustable Specimen Shelf, Horizontal Specimen Slide, Viewing Windows and Beam Hardening Filter Holder.

The TouchRAD Control Panel is a multi•user, password protected touch screen interface with a transportable database that can track individual system usage. 

Passwords, programmed exposure settings, and database management are controlled by a super-user. 

Features

  • Automatic warm-up with Intelligent Tube Conditioning
  • Proven long-term reliability of system components

Specifications

Cabinet Features
Adjustable Sample Shelf: 20cm to 90cm SSD
Changeable Beam Conditioning Filter Slides, (2) supplied
Dual Viewing Windows for specimen viewing in both vertical and horizontal configurations.
Complies with Domestic and International Regulations for safety of cabinet X-ray systems

Cabinet Size and Weight
Overall Dimensions: 86” (218cm)h x 72” (183cm)w x 40” (102cm)d
Irradiation Chamber: 42” (107cm)h x 39” (100cm)w x 28” (71cm)d
Weight: 5500 lbs (2500 kg)

Power Requirement
1N PE 230 VAC ±5%, or 3N 230 VAC ±10%, 50/60 Hz, 12.5 KVA

High Voltage Generator
Maximum Output Voltage: 450 kV (tube limited to 320 kV)
Maximum mA: 45

X-ray Tube
Maximum Potential: 320 KV
Maximum Power: 4000 W
Type: Metal Ceramic, Fixed Anode, Oil Cooled
Focal Spot: 8 mm2 (per EN12543)

Dose Output
3 Gy/min at 320KV, 12.5mA, 50cm SSD, (HVL≈1mm Cu)
1 Gy/min at 320KV, 12.5mA, 50cm SSD, (HVL≈4mm Cu)
>15 Gy/min at 320KV, 12.5mA, 50cm SSD, (No Beam Hardening))

Operators Control
KV Setting & Display Accuracy: 5KV – 320 KV in 0.1 KV increments
mA Setting & Display Accuracy: 0.5mA to 45 mA in 0.01 mA increments
Settings Accuracy: < 1%
Exposure Timer: 1-9999 seconds
Programmable Settings: >1,000 locations to recall exposure parameters
Password Protected User Accounts (> 1000 users)
Ability to assign users access to only specific programs.
Excel® Database transportable via USB Flash memory

Precision X-Ray, PXi, X-RAD 320ix, precise radiation dosage to specimens in a biological or small animal research laboratory.

Overview

The X-RAD 320ix is a self-contained X-ray system for delivering a precise radiation dosage to specimens in a biological or small animal research laboratory. 

The shielded cabinet includes a TouchRad Control, Key Operated Motorized Adjustable Specimen Shelf, Sample Viewing Window, Interior Light and Beam Hardening Filter Holder.

The X-RAD 320ix features the TouchRAD Control Panel a multi-user, password protected touch screen interface with a transportable database that can track individual system usage. 

Passwords, programmed exposure settings, and database management are controlled by a super-user.

The X-ray tube is used specifically for radiation therapy having a highly homogenous beam.

 

Features

  • Automatic warm-up with Intelligent Tube Conditioning
  • Proven long-term reliability of system components

Specifications

Cabinet Features
Adjustable Sample Shelf: 20cm to 90cm SSD
Changeable Beam Conditioning Filter Slides, (2) supplied
User Entry Port to introduce small tubing and cables into chamber area
Complies with Domestic and International Regulations for safety of cabinet X-ray systems

 

TouchRad Control
KV Setting & Display Accuracy: 5KV – 320 KV in 0.1 KV increments mA Setting & Display Accuracy: 0.5mA to 45 mA in 0.01 mA increments Settings Accuracy: < 1% Exposure Timer: 1-9999 seconds
Password Protected User Accounts (> 1000 users)
Ability to assign users access to only specific programs.
Excel® Database transportable via USB Flash memory

 

X-ray Tube
Maximum Potential: 320 KV
Maximum Power: 4000 W
Type: Metal Ceramic, Fixed Anode, Oil Cooled
Focal Spot: 8 mm2 (per EN12543)

 

Dose Output
3 Gy/min at 320KV, 12.5mA, 50cm SSD, (HVL≈1mm Cu)
1 Gy/min at 320KV, 12.5mA, 50cm SSD, (HVL≈4mm Cu)
>15 Gy/min at 320KV, 12.5mA, 50cm SSD, (No Beam Hardening))

 

High Voltage Generator
Maximum Output Voltage: 450 kV (tube limited to 320 kV)
Maximum mA: 45

Cabinet Size and Weight
Overall Dimensions: 80” (200cm) h x 59” (148cm) w x 34” (85cm) d Irradiation Chamber: 40” (100cm) h x 30” (75cm) w x 30” (75cm) d

Weight: 5500 lbs (2500 kg)

Power Requirement
1N PE 230 VAC ±5%, or 3N 230 VAC ±10%, 50/60 Hz, 12.5 KVA